Versatile applications
Optimized for the digestion of your sample materials, Speedwave Entry is suitable for use in the following areas:
Environment
Filter, soils, sediments, sludge, water, waste water
Agriculture
Grains, nutrients
Biology/medicine
tissue, blood, hair
Feed and food
ash, contaminated soils, slag
Energy
beverages, animal products, fruit, dairy products, oils
Download
Reaction control - monitoring pressure and temperature
Information brochure about sensor technology
User-friendly
Quick and easy to start: The high-quality TFMTM-PTFE digestion vessels consist of only three components and can be opened and closed manually without the need for special tools. The software offers a range of applications that are key in routine analysis.
Safe
Optimised for improved safety: if the rupture disc breaks, the integrated gas collection system safely collects all nascent gases and discharges them without risk. Thanks to the constant evacuation of the oven chamber during operation, possible emissions are effectively removed.
Cost-effective
Ideal for use in routine analysis: the compact, secure construction and ease of operation make this a flexible laboratory tool. Efficient and with high technical standards – the microwaves guarantee high-quality digestion at optimum capacity.
Pressure containers and rotor systems
Measurement
The patented visual measurement method, Speedwave® DIRC, facilitates the direct and seamless real-time recording of sample temperature in the digestion vessel. As a result, there is no risk of contamination for the samples and no risk of damage or wear and tear through the frequent assembly and disassembly of the sensors. All sensors and wiring lie outside of the microwave field and are fully chemical resistant. (See download sensor technology)
DAP-60X rotor with 10 pressure vessels
Vessel type DAP-60K
Volume 60 ml
Working pressure 580 psi
Test pressure 870 psi
Working temperature 446 °F
Rotor 10 vessels
Sample weight 300 mg (organic)
DAQ-10 with 24 pressure vessels
Vessel type DAQ-10
Volume 10 ml
Working pressure 1088 psi
Test pressure 2176 psi
Working temperature 446 °F
Rotor 24 vessels
Sample weight 100 mg (organic)